Glucose Sensor
Calibration
We
recommend allowing 3 days after implantation for the sensor to stabilize.� The 14-day sensing lifetime of the sensor is
assumed to start after this 3 day stabilization.� This time may vary based on the rat type and
individual physiology.� The procedures
described below should be viewed as guidelines, and some changes may be
appropriate for specific experimental paradigms and rat strains.
For
calibration of the CGMS biosensor, Pinnacle recommends using Two-Point Calibrations
approximately once a week along with daily single point resting glucose
measurements.� In general this
calibration strategy follows the time course below:
�
Days
3, 7, 14 � Two point calibration as described below
�
All
other days � Single point resting glucose measurement at approximately the same
time each day.� For most situations,
there will be no loss of accuracy if single point readings are not done every
day (i.e., weekends).
In
a typical scenario, the sensor will be implanted and telemetry established to
ensure a proper run-in, and the animal returned to its cage. �Our standard protocol is to perform the surgery
on Friday, so the stabilization period is complete on Monday.� The first two point calibration would then be
on Monday with additional 2 point calibrations on each subsequent Friday.
Performing a Two-Point
Calibration:
A two point calibration is typically
performed by making two separate blood glucose measurements.� One is taken when the animals
is in a quiet or resting state. This serves as the baseline measurement
and is the same as the single point resting glucose measurement.�� A second blood glucose measurement is then
taken following administration of a glucose bolus.� The glucose bolus may be administered orally,
IV or IP.� A second blood glucose
measurement should be taken after the glucose has peaked and has started its
return to baseline (see Figure 1) - approximately 10 to 15 minutes for a bolus
dose administered IP.� This timing may
change depending on the experimental model or bolus route.� Pinnacle recommends monitoring the biosensor
signal closely following bolus administration to assist in determining the
proper time-point for the second glucose measurement.
To
obtain a valid Two Point Calibration, the difference between the resting
glucose concentration and the peak of the bolus response should be greater than
100 mg/dL.� Pinnacle
recommends an IP dose of 1 g/kg glucose to achieve this difference,
however other doses and route combinations are also valid.� Blood glucose measurements may be obtained
through a variety of methods including blood vein catheterization, tail vein
bleed or tail tip cutting/bleeding.�
Blood Sampling
Technique:
For
convenience over the course of the two week experiment and to minimize animal
movement or stress while obtaining the blood sample, Pinnacle recommends using
a tail tip cleavage to obtain the blood sample.�
In this method, the tip of the animal�s tail is cut under anesthesia
during the CGMS biosensor implantation procedure.�� The scab covering this wound may then be easily
removed for daily blood glucose measurements with minimal perturbation to the
animal.
Blood Glucose Reference
Measurement:
Pinnacle
recommends using the Roche Diagnostics Accu-Chek Nano for blood glucose measurements but there are a number
of glucose monitors that will also work well.�
We also recommend that all blood glucose reference measurements be made
in duplicate, using separate test strips and monitors and be performed in quick
succession to ensure consistency.� If the
two measurements are highly divergent, a third measurement should be taken and
the measurement that is not in agreement discarded.
Entering Calibration Values
into Software:
To
store the calibration values and types directly into Sirenia
CGMS acquisition, press the �Calibration� button to evoke the panel� shown below.
Up
to 3 calibration readings can be taken at one time.� If only one is taken, leave the other fields
blank.� When the calibration values have
been entered, select the type of calibration.�
That is, point 1 of a two point, point two of a two point, or a single
point, and press �save� to enter the calibration data into the data set being
collected.� If activated, the calibration
coefficients will automatically be applied to the incoming data, so
concentration values can be displayed in real time, and alarms can be tied to
change and rate of change in mg/dl as well as nA.�
The
values entered, and additional calibration options, are available in CGMS
Analysis for detailed post processing of the data if required.